Skip to main content
  • Image

Guidion Hydro

Designed for reliable delivery of devices to target

Guidion Hydro guide extension catheter has a radiopaque soft tip for true distal end visibility facilitating precise device positioning. The high density coiling of the distal shaft ensures optimal lumen integrity and smooth device delivery. A proximal oval push rod design keeps the profile low while securing optimal pushability.1

Category

Vascular Intervention Coronary Guide Extension Catheter

Product Highlights

Image

True distal end visibility¹

For precise device positioning

Image

Flex Zoneᵃ and hydrophilic coating

For advanced trackability1

Image

Coil reinforced shaft

For smooth device delivery1

Product Overview

Guidion Hydro

Image

Technical Data

Guide Extension Catheter  
Distal shaft length 25 cm
Shaft material Polyether Block Amide
Tip Flexible, radiopaque, 1.3 mm length
Outer coating Hydrophilic distal 10 cm
Exit markers 95 cm and 105 cm
Usable length 150 cm

Size and Compatibility

BIOTRONIK/IMDS Portfolio
  Navigate Prepare Treat
Guidion Hydro Recross2 Nhancer Rx2 Pantera LEO2 Pantera Pro2 Orsiro Mission2 Pantera Lux2 Freesolve3 PK Papyrus2
5F - - - - - - - -
6F - ✔(2.5-4.0)
7F  ✔(2.5-4.0)
8F ✔(2.5-5.0)
Other Devices 
  IVUS Single Lumen Microcatheters Rotablator Burr IVL
Guidion Hydro Eagle Eye 
Platinum Family, 
Refinity ST4
ACIST 
KODAMA3
Low Profile OD
Microcatheters 
(<3F)5
Large Profile OD
Microcatheters
(>3F)5
1.25 mm6 &
1.5 mm6
Shockwave 
C27
Shockwave
C2+8
5F - - - - -
6F - -
7F  -
8F

Ordering Information

Ordering number IMDS article number Required guide catheter ID Guidion Hydro ID
451663 G50F25150 5F, ID ≥ 0.056" (1.42 mm) 0.041" (1.04 mm)
451664 G60F25150 6F, ID ≥ 0.070" (1.78 mm) 0.056" (1.42 mm)
451665 G70F25150 7F, ID ≥ 0.078" (1.98 mm) 0.062" (1.57 mm)
451666 G80F25150 8F, ID ≥ 0.088" (2.24 mm) 0.071" (1.80 mm)

How can we help you?

I prefer to be contacted by
*Fields are mandatory.

References

The compatibility conclusions provided in this document are based on general & public information. It is advised to consult the manufacturer of each device for precise compatibility details.

IVUS = Intravascular Ultrasound; IVL = Intravascular Lithotripsy

a. 6&7F. 1. IMDS data on file. 2. According to the comparison between minimum dimension of Guiding catheter of the device in the IFU and the labeled inner diameter of Guidion Hydro. 3. According to IFU. 4. According to the comparison between the requirement for Guiding Catheters (≥0.056"/ 1.42mm – Eagle Eye Platinum, Refinity ST and Eagle Eye Platinum ST, Philips IVUS Spec sheet Feb 2024) and the labeled inner diameter of Guidion Hydro. 5. Garbo R. et al. JACC Cardiovasc Interv. 2023 Aug 14;16(15):1833-1844. doi: 10.1016/j.jcin.2023.05.019. PMID: 37587590. 6. Pawłowski T. et al. Postepy Kardiol Interwencyjnej. 2021 Jun;17(2):210-213. doi: 10.5114/aic.2021.107502. Epub 2021 Jul 9. PMID: 34400924; PMCID: PMC8356840. 7. Based on C2 - PN 62185 Rev F - Shockwave C2 Coronary IVL Catheter IFU, US.pdf downloaded on Feb 7, 2024. 8. Based on PN 64192 Rev C – IFU, Shockwave C2+ Coronary IVL Catheter, US.pdf downloaded on Feb 7, 2024.

Freesolve, Lux, Orsiro, Orsiro Mission, Pantera and PK Papyrus are trademarks or registered trademarks of the BIOTRONIK Group of Companies. Guidion, NHancer Rx and ReCross are trademarks of IMDS. Manufactured by IMDS and distributed by BIOTRONIK in selected countries. All other trademarks are the property of their respective owner.